FDA Enforcement Class II Terminated

IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.

Recall: Z-1506-2016 · Reported April 27, 2016

Enforcement

Recall Number
Z-1506-2016
Event ID
73776
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 27, 2016
Initiation Date
March 4, 2016
Classification Date
April 15, 2016
Termination Date
July 2, 2018
Address
595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States

Description

IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.

Reason

The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.

Code Info

M/N 881001; S/N: 85254, 82202, 85330, 82117, 7000022, 105009, 970001, 850001, 870030, 82141, 95068, 87182, 87197, 95078, 82140, 11017, 98139, 95035, 87200, 980047, 85314, 85303, 980046, 85298, 822022, 7000019, 950076, 950115, 95043, 95047, 950021, 960123 & 105006.

Distribution

Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.

Quantity

33 Units