ADVIA Centaur Cortisol 50T (Material Number 10994924), ADVIA Centaur Cortisol 250T (Material Number 10994926), and ADVIA Centaur Cortisol REF 250T (Material Number 10994927). In vitro diagnostic use in the quantitative determination of cortisol in human serum, plasma (EDTA and lithium heparin), and urine.
Enforcement
- Recall Number
- Z-1505-2023
- Event ID
- 92059
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 10, 2023
- Initiation Date
- March 16, 2023
- Classification Date
- May 3, 2023
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005, United States
Description
ADVIA Centaur Cortisol 50T (Material Number 10994924), ADVIA Centaur Cortisol 250T (Material Number 10994926), and ADVIA Centaur Cortisol REF 250T (Material Number 10994927). In vitro diagnostic use in the quantitative determination of cortisol in human serum, plasma (EDTA and lithium heparin), and urine.
There is a negative bias with urine patient samples and urine cortisol Quality Control (QC) results intermittently out of range low.
ADVIA Centaur Cortisol (50T) UDI-DI: 00630414602943; Kit Lot Numbers: 49741344 07241346 88318346 29109348 and higher; ADVIA Centaur Cortisol 250T UDI-DI: 00630414602950; Kit Lot Numbers: 49742344 88319346 29108348 and higher; ADVIA Centaur Cortisol (REF) 250T UDI-DI: 00630414598659; Kit Lot Numbers: 75474344 11352346 and higher
Worldwide distribution - US Nationwide.
18,961 kits