FDA Enforcement
Class II
Terminated
Specialty MiniScope Mech. Set Size 18, Specialty MiniScope Mech. Set Size 21, and Specialty MiniScope Mech. Set Size 24 A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.
Recall: Z-1505-2015
·
Reported April 29, 2015
Enforcement
- Recall Number
- Z-1505-2015
- Event ID
- 70787
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Specialty Appliance Works, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 29, 2015
- Initiation Date
- March 9, 2015
- Classification Date
- April 22, 2015
- Termination Date
- January 12, 2017
- Address
- 4905 Hammond Industrial Dr, Cumming, GA, 30041-3914, United States
Description
Specialty MiniScope Mech. Set Size 18, Specialty MiniScope Mech. Set Size 21, and Specialty MiniScope Mech. Set Size 24 A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.
Reason
The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.
Code Info
Product No. 90870, 90876 and 90879
Distribution
Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.
Quantity
14,428 Universal Nuts