FDA Enforcement Class II Terminated

Giraffe OmniBed Ohmeda Medical Giraffe OmniBed- The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperatures controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Recall: Z-1504-2017 · Reported March 29, 2017

Enforcement

Recall Number
Z-1504-2017
Event ID
76283
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ohmeda Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 29, 2017
Initiation Date
November 18, 2016
Classification Date
March 20, 2017
Termination Date
July 13, 2018
Address
8880 Gorman Rd, Laurel, MD, 20723-5800, United States

Description

Giraffe OmniBed Ohmeda Medical Giraffe OmniBed- The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperatures controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Reason

Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.

Code Info

See Attached Spreadsheat

Distribution

Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,

Quantity

5,993 (5,080 US; 913 OUS)