FDA Enforcement
Class II
Ongoing
Ultraview SL (UVSL) Command Module, Model 91496
Recall: Z-1503-2023
·
Reported May 10, 2023
Enforcement
- Recall Number
- Z-1503-2023
- Event ID
- 91851
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Spacelabs Healthcare, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 10, 2023
- Initiation Date
- April 12, 2023
- Classification Date
- May 3, 2023
- Address
- 35301 Se Center St, Snoqualmie, WA, 98065-9216, United States
Description
Ultraview SL (UVSL) Command Module, Model 91496
Reason
Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
Code Info
Software version 2.03.06 and earlier and software version 2.04 and later.
Distribution
US Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR, TX, GA, IL, OK, SC, PA, AL, KS, FL, ID, MI, MN, IA, NJ, MD, VA, AR, CT, NE, LA, AK, MS, ME, WV, NY, AZ, TN, WY, IN, MA, MT, CO, DE, ND, WI, NV, RI, NH, UT.
Quantity
1,088