AGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Usage: TomoSynthesis is used to synthesize tomographic slices from a single tomographic sweep. The DR 800 is not intended for mammography applications
Recall: Z-1502-2020
·
Reported March 25, 2020
Enforcement
- Recall Number
- Z-1502-2020
- Event ID
- 85028
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Agfa N.V.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- March 25, 2020
- Initiation Date
- February 10, 2020
- Classification Date
- March 18, 2020
- Termination Date
- March 30, 2023
- Address
- Septestraat 27, Mortsel, N/A, Belgium
Description
AGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Usage: TomoSynthesis is used to synthesize tomographic slices from a single tomographic sweep. The DR 800 is not intended for mammography applications
Reason
Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.
Code Info
Serial Numbers: 00949, 1161
Distribution
US Nationwide distribution in the states of FL, VA.
Quantity
2 units