FDA Enforcement Class II Terminated

Ysio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.

Recall: Z-1501-2020 · Reported March 25, 2020

Enforcement

Recall Number
Z-1501-2020
Event ID
85029
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 25, 2020
Initiation Date
February 10, 2020
Classification Date
March 16, 2020
Termination Date
August 31, 2020
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

Ysio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.

Reason

The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the systems controls may (depending on the values being processed) initiate a correction of the tube stand position, resulting in an unexpected movement.

Code Info

Serial Nos. 20113 20034 21100 21086 21113 21088 21103 21221 21223 21224 21085 20045 20093 20110 21066 21031 21139 21006 20155 21112 21122 21185 20088 20089 20106 20107 21052 21089 21213 20167 21098 21174 21057 21101 21116 21117 21118 21120 21091 21092 21037 21076 21077 21200 21203 21209 20149 21163 21172 21171 21173 20094 21115 20047 21194 20193 21211 21212 21186 21188 21094 21039 21040 21108 20035 20185 21061 21065 21205 21095 21148 21189 21140 21175 21023 21001 21127 21123 21013 20159 21160

Distribution

US Nationwide distribution in the states of AR, CA, DC, FL, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, UT, VA and WI.

Quantity

81 US/ 308 worldwide