Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examination. It increases ambient light to facilitate performance of procedures on the mattress field. It is designed to mount on the accessory railings of the Giraffe OmniBed, incubators and warmers except for Panda and Giraffe Warmers. It is not intended for use to support surgical field lighting.
Enforcement
- Recall Number
- Z-1501-2017
- Event ID
- 76283
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ohmeda Medical
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- March 29, 2017
- Initiation Date
- November 18, 2016
- Classification Date
- March 20, 2017
- Termination Date
- July 13, 2018
- Address
- 8880 Gorman Rd, Laurel, MD, 20723-5800, United States
Description
Giraffe Exam Light The Exam Light illuminates body surfaces and facilitates patient observation during a medical examination. It increases ambient light to facilitate performance of procedures on the mattress field. It is designed to mount on the accessory railings of the Giraffe OmniBed, incubators and warmers except for Panda and Giraffe Warmers. It is not intended for use to support surgical field lighting.
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
See Attached Spreadsheat
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
5,993 (5,080 US; 913 OUS)