iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA only) and 2016.1.0.127 (out of US) Opthalmic: iVue is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber the eye.
Enforcement
- Recall Number
- Z-1498-2016
- Event ID
- 73612
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Optovue, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 27, 2016
- Initiation Date
- March 16, 2016
- Classification Date
- April 15, 2016
- Termination Date
- April 4, 2017
- Address
- 2800 Bayview Dr, N/A, Fremont, CA, 94538-6518, United States
Description
iVue and iVue 500 (now called iScan) OCT systems with software version 2016.0.0.127 (USA only) and 2016.1.0.127 (out of US) Opthalmic: iVue is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (GCC), optic disc, cornea, and anterior chamber the eye.
A software anomaly was detected that affects Optovue iVue and iVue 500 with Normative Database that may display images of the same eye rather than right and left if user scrolls on the map.
US and foreign (not yet provided) Software version 2016.0.0.127: Serial numbers: 21908, 26507, 26496, 26558, 22738, 26139, 26442, 25720, 25079, 25142, 23298, 21990, 23461, 24199, 25843, 26572, 26550, 22971, 26495, 22652, 26550, 22971, 26495, 22652, 25389, 23309, 23760, 21732-1, 23527, 20923, 21600, 21414, 20271, 23153, 22875, 23211. Software version 2016.1.0.127; Serial numbers forthcoming.
PA, CA, IL, HI. MO, CO, GA, ME, IN, AZ, WV, VA, FL, AR, UT, NE, IN, NV, LA, TN, KS, OK. Foreign distribution forthcoming
40