FDA Enforcement
Class II
Ongoing
IBA Proton Therapy System - PROTEUS 235
Recall: Z-1497-2026
·
Reported March 18, 2026
Enforcement
- Recall Number
- Z-1497-2026
- Event ID
- 98364
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Ion Beam Applications S.A.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 18, 2026
- Initiation Date
- February 3, 2026
- Classification Date
- March 6, 2026
- Address
- Chemin Du Cyclotron 3, N/A, Ottignies-Louvain-La-Neuve, N/A, N/A, Belgium
Description
IBA Proton Therapy System - PROTEUS 235
Reason
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Code Info
PROTEUS 235; UDI-DI: 05404013801138; Serial Numbers: SBF103 (JP), SBF124 (EU), SBF133 (US), SBF140 (EU)
Distribution
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
Quantity
4 units (1 US, 3 OUS)