Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.
Enforcement
- Recall Number
- Z-1497-2019
- Event ID
- 82595
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Abbott Gmbh & Co. KG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 5, 2019
- Initiation Date
- April 5, 2019
- Classification Date
- May 30, 2019
- Termination Date
- September 28, 2022
- Address
- Max-Planck-Ring 2, 65205, Wiesbaden, N/A, N/A, Germany
Description
Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.
Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.
Part Number: 03R65 01, all Serial Numbers; a. TED (Thermo Electric Device) Engine, Reagent cooler - Part Numbers A 30103732 01, A 30103732 02 with all Serial Numbers; b. CABLE, POWER OUT, TED CNTRLER - Part Numbers A 35006203 01, A 35006203 02 with all Serial Numbers
Worldwide Distribution - US Nationwide to Oklahoma and Texas. Distributed worldwide to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, MADAGASCAR, MEXICO, NETHERLANDS, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UAE, UNITED KINGDOM, VIETNAM.
373 units total