FDA Enforcement Class I Ongoing

ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814RM

Recall: Z-1495-2023 · Reported May 17, 2023

Enforcement

Recall Number
Z-1495-2023
Event ID
91922
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Cordis US Corp
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 17, 2023
Initiation Date
March 24, 2023
Classification Date
May 11, 2023
Address
14201 Nw 60th Ave, N/A, Miami Lakes, FL, 33014-2802, United States

Description

ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814RM

Reason

There is a potential for separation of the ANGIOGUARD RX / XP delivery system and capture sheath which include but are not limited to situations of an intra-procedural delay, unplanned percutaneous or surgical intervention, or stroke; while a replacement device is prepared.

Code Info

UDI/DI 20705032054796; Lot Numbers: 35265644

Distribution

Worldwide distribution.

Quantity

N/A