FDA Enforcement Class II Terminated

Stryker Instruments 0400-770-000 T4 Toga, Pullover, Ex-Large Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

Recall: Z-1494-2013 · Reported June 12, 2013

Enforcement

Recall Number
Z-1494-2013
Event ID
65321
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Instruments Div. of Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
June 12, 2013
Initiation Date
May 23, 2013
Classification Date
June 6, 2013
Termination Date
April 14, 2014
Address
4100 E Milham Ave, N/A, Portage, MI, 49002-9704, United States

Description

Stryker Instruments 0400-770-000 T4 Toga, Pullover, Ex-Large Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

Reason

Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

Code Info

1005105, 1009214, 1104308, 1006130, 1010260, 1104312, 1007152, 1012346, 1105378, 1007163, 1012303, 1106460, 1007164, 1101062, 1107542, 1008189, 1101063, 1108660, 1008188, 1102149, 1109760, 1009213, and 1103219.

Distribution

USA Nationwide Distribution

Quantity

21,660 units (2166 boxes)