FDA Enforcement Class II Ongoing

Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754

Recall: Z-1492-2025 · Reported April 9, 2025

Enforcement

Recall Number
Z-1492-2025
Event ID
96331
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Terumo Cardiovascular Systems Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 9, 2025
Initiation Date
March 3, 2025
Classification Date
March 31, 2025
Address
6200 Jackson Rd, Ann Arbor, MI, 48103-9586, United States

Description

Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754

Reason

During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.

Code Info

UDI/DI 00886799001882, Serial Number range: H0050095-H0050110, H0050112-H0050116, H0050118-H0050124, H0050126, H0050131-H0050134, H0050136, H0050140-H0050142

Distribution

US: IL, TX, NY, MI, FL, NV, and Australia, New Zealand

Quantity

37 units