FDA Enforcement
Class II
Ongoing
Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
Recall: Z-1492-2025
·
Reported April 9, 2025
Enforcement
- Recall Number
- Z-1492-2025
- Event ID
- 96331
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Terumo Cardiovascular Systems Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 9, 2025
- Initiation Date
- March 3, 2025
- Classification Date
- March 31, 2025
- Address
- 6200 Jackson Rd, Ann Arbor, MI, 48103-9586, United States
Description
Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754
Reason
During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.
Code Info
UDI/DI 00886799001882, Serial Number range: H0050095-H0050110, H0050112-H0050116, H0050118-H0050124, H0050126, H0050131-H0050134, H0050136, H0050140-H0050142
Distribution
US: IL, TX, NY, MI, FL, NV, and Australia, New Zealand
Quantity
37 units