Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116. Proton therapy

Recall: Z-1492-2014 · Reported April 30, 2014

Enforcement

Recall Number
Z-1492-2014
Event ID
67678
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ion Beam Applications S.A.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 30, 2014
Initiation Date
December 4, 2013
Classification Date
April 24, 2014
Termination Date
June 12, 2014
Address
Chemin du Cyclotron, 3, N/A, Louvain La Neuve, N/A, N/A, Belgium

Description

Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116. Proton therapy

Reason

Internal testing of the Proteus 235 Proton Therapy System revealed incorrect dose rate calculation in case of beam line option other than Zero (0).

Code Info

Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116.

Distribution

US Distribution including the states of OK, IL, NJ, and WA.

Quantity

4