FDA Enforcement
Class II
Terminated
Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116. Proton therapy
Recall: Z-1492-2014
·
Reported April 30, 2014
Enforcement
- Recall Number
- Z-1492-2014
- Event ID
- 67678
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ion Beam Applications S.A.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 30, 2014
- Initiation Date
- December 4, 2013
- Classification Date
- April 24, 2014
- Termination Date
- June 12, 2014
- Address
- Chemin du Cyclotron, 3, N/A, Louvain La Neuve, N/A, N/A, Belgium
Description
Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116. Proton therapy
Reason
Internal testing of the Proteus 235 Proton Therapy System revealed incorrect dose rate calculation in case of beam line option other than Zero (0).
Code Info
Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116.
Distribution
US Distribution including the states of OK, IL, NJ, and WA.
Quantity
4