VITROS Chemistry Products CDM PROM (Calibration Data Module) No. 0168, Catalog Number/REF 199 9077, used in conjunction with: VITROS Chemistry Products CRSC DT Slides, Generation (GEN) 82, REF 6802721, and VITROS DT Calibrator Kit, Product Code 1957924, and tested on the VITROS DT60/DT60 II Chemistry System; IVD --- Ortho-Clinical Diagnostics, Inc.
Enforcement
- Recall Number
- Z-1490-2014
- Event ID
- 67509
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ortho-Clinical Diagnostics
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 30, 2014
- Initiation Date
- January 29, 2014
- Classification Date
- April 23, 2014
- Termination Date
- May 18, 2018
- Address
- 1000 Lee Road, N/A, Rochester, NY, 14606, United States
Description
VITROS Chemistry Products CDM PROM (Calibration Data Module) No. 0168, Catalog Number/REF 199 9077, used in conjunction with: VITROS Chemistry Products CRSC DT Slides, Generation (GEN) 82, REF 6802721, and VITROS DT Calibrator Kit, Product Code 1957924, and tested on the VITROS DT60/DT60 II Chemistry System; IVD --- Ortho-Clinical Diagnostics, Inc.
Customers were unable to calibrate, or use previously calibrated, VITROS DT CRSC slides (REF/Product Code 6802721), GEN 82, on the VITROS DT II System, when using Calibration Data Module (CDM) Rev. 168 (REF/Product Code 199 9077), as CDM 168 does not contain calibration parameters for DT CRSC.
Calibration Data Module Rev. 0168, REF 199 9077 used in conjuction with of VITROS Chemistry Products CRSC DT Slides, Generation (GEN) 82, Product Code 6802721.
Worldwide Distribution - US Distribution including the states of AR, FL, IA, IL, KY, MI, MN, ND, NY and VA, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
US: 284 units, Ex-US: 4772 units