FDA Enforcement Class II Terminated

VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor adjustment procedure.

Recall: Z-1489-2021 · Reported May 5, 2021

Enforcement

Recall Number
Z-1489-2021
Event ID
87649
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, INc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 5, 2021
Initiation Date
March 26, 2021
Classification Date
April 23, 2021
Termination Date
May 24, 2023
Address
513 Technology Blvd, N/A, Rochester, NY, 14626-3601, United States

Description

VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor adjustment procedure.

Reason

Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesnt have any White Reference Slides in reserve, they cant run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.

Code Info

Lots: 9252-0058-3361, 9252-0058-3362

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV, WY and the countries of Australia, Bermuda, Brazil, Canada, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.

Quantity

865 units of 25 slides each