FDA Enforcement
Class II
Terminated
Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
Recall: Z-1487-2020
·
Reported March 25, 2020
Enforcement
- Recall Number
- Z-1487-2020
- Event ID
- 84756
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips North America, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 25, 2020
- Initiation Date
- December 30, 2019
- Classification Date
- March 13, 2020
- Termination Date
- November 17, 2022
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032, United States
Description
Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
Reason
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Code Info
Serial Codes 3217017 200 58 72 CV.170 180 524615 47 211 364 30 3862570/0011 529 S.N. 519 S01H060871 505/0050114 S01H057274 01H077213 530 532 S01HA52566 348 476 320 481 467 284 352 359 459 73*004 733 536 CV.474 454 49 281 51 186 122 250 37 22 CV.312 296 373 CV.596 176 30 580 279
Distribution
Nationwide domestic distribution. Foreign distribution worldwide.
Quantity
70