FDA Enforcement
Class II
Terminated
iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.
Recall: Z-1487-2014
·
Reported April 30, 2014
Enforcement
- Recall Number
- Z-1487-2014
- Event ID
- 67959
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Boston Scientific Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 30, 2014
- Initiation Date
- April 3, 2014
- Classification Date
- April 22, 2014
- Termination Date
- March 18, 2015
- Address
- 47215 Lakeview Blvd, N/A, Fremont, CA, 94538-6530, United States
Description
iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.
Reason
Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.
Code Info
Model Numbers: H749ILAB120CART0 H749ILAB120INS0 H749ILAB120N2710 Serial numbers: 4219, 4900, 5701, 5827 7516, 7517, 7607 8407, 8434. the different UPNs represent minor level differences in configuration relative to mobility and no difference in any essential components or software.
Distribution
USA Nationwide Distribution in the states of NY, NC, DC, FL, PA
Quantity
9 units