FDA Enforcement
Class II
Ongoing
KWIK-STIK Rhizopus stolonifer (+) derived from ATCC 6227b, a) REF 0209P and b) REF 0209K, Quality control kit for culture media
Recall: Z-1483-2024
·
Reported April 17, 2024
Enforcement
- Recall Number
- Z-1483-2024
- Event ID
- 94201
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Microbiologics Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 17, 2024
- Initiation Date
- March 6, 2024
- Classification Date
- April 5, 2024
- Address
- 200 Cooper Ave N, N/A, Saint Cloud, MN, 56303-4440, United States
Description
KWIK-STIK Rhizopus stolonifer (+) derived from ATCC 6227b, a) REF 0209P and b) REF 0209K, Quality control kit for culture media
Reason
The positive control material (Rhizopus stolonifer) within the KWIK-STIK assembly was contaminated with another fungal organism (Aspergillus brasiliensis). As a result, users may observe A. brasiliensis growth when using the control material.
Code Info
a) REF 0209P, UDI/DI 20845357003113, Lot Numbers: 209-37-1, 209-37-3, 209-37-4, 209-37-5; b) REF 0209K, UDI/DI 30845357003127, Lot Numbers: 209-37-2
Distribution
Worldwide - US Nationwide distribution in the states of CA, MI, KY, MO, NM, TN, PR, and the countries of France, Columbia, Kazakhstan, Peru, Puerto Rico, South Africa.
Quantity
19 units