FDA Enforcement
Class II
Terminated
Straight Impactor - Dorado IBC, Part Number: SI02140. The device is used to position implants after they have been inserted into the disc space during implantation of intervertebral body fusion devices.
Recall: Z-1482-2013
·
Reported June 12, 2013
Enforcement
- Recall Number
- Z-1482-2013
- Event ID
- 64991
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- SpineFrontier, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Other
- Report Date
- June 12, 2013
- Initiation Date
- May 10, 2008
- Classification Date
- June 5, 2013
- Termination Date
- October 8, 2014
- Address
- 500 Cummings Ctr Ste 3500, N/A, Beverly, MA, 01915-6516, United States
Description
Straight Impactor - Dorado IBC, Part Number: SI02140. The device is used to position implants after they have been inserted into the disc space during implantation of intervertebral body fusion devices.
Reason
During two separate occasions an impactor plate on the handle broke off while being struck by a mallet.
Code Info
Lot number: 20856
Distribution
Worldwide distribution: US (nationwide) to states of: CO, MA, and TX; and country of: Jamaica.
Quantity
10