FDA Enforcement
Class II
Ongoing
EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
Recall: Z-1480-2026
·
Reported March 11, 2026
Enforcement
- Recall Number
- Z-1480-2026
- Event ID
- 98397
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Staar Surgical AG
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- N/A
- Report Date
- March 11, 2026
- Initiation Date
- January 21, 2026
- Classification Date
- March 2, 2026
- Address
- Haupt Strasse 104, N/A, Nidau, N/A, N/A, Switzerland
Description
EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5
Reason
Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.
Code Info
UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988
Distribution
International distribution to the countries of India, Iran, Korea, Saudi Arabia.
Quantity
7