FDA Enforcement Class II Ongoing

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Recall: Z-1480-2026 · Reported March 11, 2026

Enforcement

Recall Number
Z-1480-2026
Event ID
98397
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Staar Surgical AG
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
March 11, 2026
Initiation Date
January 21, 2026
Classification Date
March 2, 2026
Address
Haupt Strasse 104, N/A, Nidau, N/A, N/A, Switzerland

Description

EVO+ Visian Toric Implantable Collamer Lens (TICL), Model: VTICM5

Reason

Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant shallowing of anterior chamber, Patient discomfort, Elevated IOP/Angle closure, Unreactive pupil, Lens exchange/removal.

Code Info

UDI-DI: 00840311304363. Expiration Date: 2027-11-30 Serial Numbers: T2286964, T2286971, T2286973, T2286979, T2286980, T2286981, T2286988

Distribution

International distribution to the countries of India, Iran, Korea, Saudi Arabia.

Quantity

7