882480: BrightView designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar static, dynamic, gated, total body, circular-orbit and noncircular orbit SPECT, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors).
Enforcement
- Recall Number
- Z-1480-2017
- Event ID
- 76483
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Medical Systems (Cleveland) Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- March 22, 2017
- Initiation Date
- January 31, 2017
- Classification Date
- March 16, 2017
- Termination Date
- July 2, 2018
- Address
- 595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States
Description
882480: BrightView designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar static, dynamic, gated, total body, circular-orbit and noncircular orbit SPECT, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors).
Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan
4000201, 40007030003, 40007060006, 4000586, 4000327, 4000307, 4000361, 4000146, 4000159, 4000077, 4000050, 4000687, 4000325, 4000165, 4000252, 4000315, 4000191, 4000598, 4000141, 4000565, 4000135, 4000537, 4000591, 4000311, 4000626, 4000636, 4000329, 4000181, 4000644, 4000171, 4000120, 4000484A, 4000441, 4000054, 4000110, 4000653, 4000590, 4000100, 4000520, 4000323, 4000403, 4000406, 4000489, 4000277, 4000306, 4000332, 4000635, 4000270, 4000302, 4000046, 4000214, 4000615, 4000187, 4000660, 4000383, 4000633, 4000322, 4000212, 4000628, 4000034, 4000232, 4000668, 4000480, 4000430, 4000553, 4000467, 4000221, 4000127, 4000259, 4000670, 4000193, 4000222, 4000289, 4000597, 4000439, 4000542, 4000531, 4000588, 4000358, 4000330, 4000508, 4000310, 4000371, 4000501, 4000609, 4000616, 4000116, 4000452, 4000058, 4000521, 4000059, 4000064, 4000482, 4000384, 4000528, 4000177, 4000506, 4000278, 4000089, 4000139, 4000130, 4000564, 4000578, 4000674, 4000083, 4000583, 4000225, 4000267, 4000563, 4000587, 4000117, 4000053, 4000662, 4000526, 4000207, 4000118, 4000103, 4000556, 4000247, 4000326, 4000174, 4000248, 4000642, 4000696, 4000182, 4000456, 4000060, 4000522, 4000448, 4000099, 4000246, 4000175, 4000524, 4000630, 4000155, 4000367, 4000360, 4000492, 4000037, 4000230, 4000308, 4000236, 4000143, 4000285, 4000505, 4000206, 4000684, 4000164, 4000004, 4000650, 4000671, 4000473, 4000405, 4000235, 4000242, 4000680, 4000318, 4000561, 4000149, 4000073, 4000240, 4000606, 4000610, 4000304, 4000584, 4000580, 4000460, 4000263, 4000347, 4000632, 4000265, 4000368, 4000055, 4000349, 4000658, 4000351, 4000388, 4000589, 4000558, 4000106, 4000144, 4000444, 4000292, 4000098A, 4000319, 4000245, 4000295, 400054, 4000470, 4000666, 4000111, 4000148, 4000076, 4000443, 4000575, 4000654, 4000251, 4000108, 4000657, 4000659, 4000562, 4000535, 4000435, 4000357,4000677, 4000453, 4000377, 4000387, 4000446, 4000608, 40007060008, 4000036, 4000433, 4000538, 4000474, 4000414, 4000569, 4000266, 4000204, 4000119, 4000079, 4000234, 4000253, 4000192, 4000249, 4000634, 4000152, 4000380, 4000513, 4000198, 4000313, 4000340, 4000525, 4000541, 4000231, 4000424, 4000161, 4000170, 4000048, 4000239, 4000124, 4000200, 40007050005, 40007060007, 4000625, 4000264, 4000534, 4000261, 4000359, 4000220, 4000136, 4000140, 4000355, 4000087, 4000104, 4000611, 4000002, 4000040, 4000093, 4000258, 4000352, 4000291, 4000132, 4000334, 4000486, 4000271, 4000599, 4000605, 4000196, 4000363, 4000656, 4000362, 4000074, 4000577, 4000151-1, 4000238, 4000007, 4000113, 4000284, 4000260, 4000244, 4000391, 4000344, 4000695, 4000112, 4000205, 4000250, 4000052, 4000432, 4000275, 4000185, 4000646, 4000094, 4000215, 4000507, 4000090, 40000101, 4000582, 4000594, 4000128, 4000661, 4000431, 4000703001, 4000523, 4000078, 4000425, 4000062, 4000145, 4000011, 4000426, 4000440, 4000529, 4000450, 4000509, 4000241, 4000320, 40000092, 4000603, 4000496, 4000532, 4000502, 4000255, 4000593, 4000490, 4000293, 4000434, 4000619, 4000438, 4000095, 4000499, 4000623, 4000517, 4000624, 4000162, 4000402, 4000574, 4000476, 4000629, 4000649, 4000685, 4000596, 4000160, 4000546, 4000622, 4000186, 4000189, 4000445, 4000429, 4000274, 4000697, 4000673, 4000038, 4000176, 4000404, 4000114, 4000309, 4000138, 4000188, 4000369, 4000178, 4000459, 4000461, 4000468, 4000063, 4000150, 4000210, 4000217, 4000547, 4000527, 4000279, 4000208, 4000216, 4000269, 4000125, 4000254, 4000345, 4000415, 4000282, 4000449, 4000679, 4000199, 4000211, 4000331, 4000082, 4000134, 4000163, 4000373, 4000409, 4000465, 4000471, 4000545, 4000618, 4000637, 4000641, 4000153, 4000539, 4000663, 4000676, 4000477, 4000299, 4000133, 4000675, 4000314, 4000686, 4000209, 4000010, 4000621, 4000041, 4000167, 4000203, 4000407, 4000543, 4000374, 4000428, 4000157, 4000013, 4000075, 4000080, 4000086, 4000392, 4000396, 4000481, 4000689, 4000420, 4000690, 4000530, 4000639, 4000479, 4000568, 4000300, 4000572, 4000573, 4000640, 4000183, 4000142, 4000066, 4000226, 4000475, 4000519, 4000691, 4000645, 4000035, 4000219, 4000485, 4000643, 4000376, 4000416, 4000243, 4000437, 4000694, 4000551, 4340004, 4000272, 4000478, 4000296, 4000354, 4000493, 4000072, 4000237, 4000107, 4000227, 4000105, 4000698, 4000333, 4000339, 4000385, 4000109, 4000115, 4000356, 4000257, 4000173, 4000126, 4000121, 4000494, 4000398, 4000503, 4000386, 4000399, 4000179, 4000600, 4000400, 4000397, 4000194, 4000154, 4000504, 4000233, 4000256, 4000667, 4000401, 4000394, 4000585, 4000168, 4000631, 4000268, 4000382, 4000692, 4000693, 4000337, 4000595, 4000447, 4000273, 4000169, 4000442, 4000412, 4000488, 4000297, 4000228, 4000156, 4000047, 4000049, 4000096, 4000281, 4000288, 4000484, 4340005, 10422, 4000419, 4000462, 4000463, 4000464, 4000638, 4000390, 4000348, 4000613, 4000655, 4000286, 4000294, 4000395, 4000672, 4000057-OLD, 4000389, 4000290, 4000283, 4000312, 4000328, 4000375, 4000560, 4000366, 4000417, 4000051, 4000372, 4000151, 4000422, 4000195, 4000088, 4000381, 4000071, 4000003, 4000549, 4000487, 4000056, 4000006, 4000466, 4000550, 4000346, 4000683, 4000483, 4000301, 4000158, 4000223, 4000045, 4000129, 4000057, 4000557, 4000567, 4000370, 4000536, 4000029, 4000607, 4000067, 4000068, 4000032, 4000423, 4000039, 4000031, 4000620, 4000678, 4000601, 4000122, 4000647, 4000341, 4000316, 4000342, 4000408, 4000421, 4000495, 4000511, 4000664, 4000317, 4000338, 4000604, 4000682, 4000540, 4000137, 4000343, 4000030, 4000379, 4000451, 4000533, 4000602, 4000472A, 4000581, 4000469, 4000648, 4000510, 4000454, 4000335, 4000436, 4000353, 4000418, 4000491, 4000571, 4000197, 4000555, 4000518, 4000544, 4000500, 4000592, 4000617, 4000012, 4000213, 4000413, 4000699, 4000218, 4000123, 4000287, 4000070, 4000166, 4000202, 4000102, 4000276, 4000091, 4000301-1, 4000455, 4000033, 4000364, 4000008, 4000497, 4000498, 4000681, 4000009, 4000184, 4000554, 4000516, 4000303, 4000576, 4000190, 4000305, 4000411, 4000669, 4000614, 4000324, 4000612, 4000566, 4000627, 4000552, 4000280, 4000378, 4000336, 4000559, 4000229, 4000262, 4000298, 4000043, 400000275, 4000044, 4000065, 4000061, 4000180, 4000224, 4000321, 4000081, 4000410, 4000651, 4000652, 4000665, 4000688 & 5000916,
Worldwide distribution. US nationwide, ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CAYMEN ISLANDS, CHILE, CHINA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, EGYPT, FINLAND, FRANCE, GABON, GERMANY, GREAT BRITAIN, GREECE, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, LEBANON, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, & VIETNAM.
1218 total