FDA Enforcement Class II Ongoing

Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITL

Recall: Z-1479-2025 · Reported April 9, 2025

Enforcement

Recall Number
Z-1479-2025
Event ID
96393
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 9, 2025
Initiation Date
February 25, 2025
Classification Date
March 28, 2025
Address
1941 Stryker Way, Portage, MI, 49002-9711, United States

Description

Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITL

Reason

There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.

Code Info

5400-030-008 UDI-DI: 07613154340842, Lot Numbers: 2023012515 2023012715 2023022012 2023050814 2023072632 2023102312 2023102612 2024032514 2024040212; 5400-030-008ITL UDI-DI: 07613327359473, Lot Numbers: 2022092714 2023020633 2023032433 2023050915 2023052514 2023062216 2023082814 2023101112 2024031812 2022092914 2023020814 2023032814 2023051114 2023053115 2023062032 2023090618 2023101714 2024042212 2022101332 2023020915 2023033016 2023051215 2023060914 2023082215 2023100312 2023102014 2024042312 2022121932 2023021416 2023050115 2023052214 2023061412 2023082312 2023100412 2024030414 2024061814 2022122032 2023022032 2023050515 2023052412 2023061514 2023082514 2023100912 2024030512 2024062814

Distribution

US Nationwide distribution.

Quantity

182,344 total units