HIGH PRESSURE ROTATOR (Component) a. WITH MALE LUER LOCK, Model Number MX496HP b. WITH FEMALE LUER LOCK, Model Number MX497HP. component of extravascular blood-pressure Transducer.
Enforcement
- Recall Number
- Z-1479-2021
- Event ID
- 87647
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Smiths Medical ASD Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 28, 2021
- Initiation Date
- March 31, 2021
- Classification Date
- April 21, 2021
- Termination Date
- March 20, 2023
- Address
- 6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States
Description
HIGH PRESSURE ROTATOR (Component) a. WITH MALE LUER LOCK, Model Number MX496HP b. WITH FEMALE LUER LOCK, Model Number MX497HP. component of extravascular blood-pressure Transducer.
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.
a. Model Number MX496HP, Lot Numbers: 3952480, 3974771 b. Model Number MX497HP, Lot Numbers: 3949391, 3971003
Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.
8150 units