FDA Enforcement Class II Ongoing

NaviNetics Skull Anchor Key and Drill Kit, REF NN1215

Recall: Z-1478-2024 · Reported April 10, 2024

Enforcement

Recall Number
Z-1478-2024
Event ID
94163
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Navinetics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 10, 2024
Initiation Date
February 26, 2024
Classification Date
April 4, 2024
Address
206 S Broadway Ste 700, Rochester, MN, 55904, United States

Description

NaviNetics Skull Anchor Key and Drill Kit, REF NN1215

Reason

Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.

Code Info

UDI/DI 10850023620276, Lot Number 331401724

Distribution

US Nationwide distribution in the states of FL, MN, CO.

Quantity