Stopcock rotator (Component): a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model Number MX1431MRL c. 700 PSI STPCK W/ ROTATOR, Model Number MX4331R. Stopcocks and manifolds used to control the direction of IV fluid flow.
Enforcement
- Recall Number
- Z-1477-2021
- Event ID
- 87647
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Smiths Medical ASD Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 28, 2021
- Initiation Date
- March 31, 2021
- Classification Date
- April 21, 2021
- Termination Date
- March 20, 2023
- Address
- 6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States
Description
Stopcock rotator (Component): a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model Number MX1431MRL c. 700 PSI STPCK W/ ROTATOR, Model Number MX4331R. Stopcocks and manifolds used to control the direction of IV fluid flow.
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.
a. Model Number MX1431MR, Lot Numbers: 3995544 b. Model Number MX1431MRL, Lot Numbers: 3995545, 4011838, 4027758, 4033501, 4044959 c. Model Number MX4331R, Lot Numbers: 3971467, 3987918
Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.
3825 units