FDA Enforcement Class II Terminated

C2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller; Cryosurgical Unit, Cryogenic Surgical Device The affected product is the Controller Cap component only, which is provided to the customer already assembled onto the Controller. Model number: FG-1012 Product Usage: The C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett s Esophagus with dysplasia. It is a cryosurgical system with a nitrous oxide cooled balloon that is compatible with a commercially available endoscope with a minimum working channel inner diameter (ID) of 3.7 mm and length of 100 cm. The C2 CryoBalloon System is a system comprised of a Catheter (sterile), Controller (non-sterile), and Cartridge (non-sterile). The Device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the inflated balloon, and the treatment site is selected by adjusting the endoscope and Controller position. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.

Recall: Z-1477-2017 · Reported March 22, 2017

Enforcement

Recall Number
Z-1477-2017
Event ID
76358
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
C2 Therapeutics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
March 22, 2017
Initiation Date
January 6, 2017
Classification Date
March 15, 2017
Termination Date
September 13, 2017
Address
303 Convention Way Ste 1, N/A, Redwood City, CA, 94063-1415, United States

Description

C2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller; Cryosurgical Unit, Cryogenic Surgical Device The affected product is the Controller Cap component only, which is provided to the customer already assembled onto the Controller. Model number: FG-1012 Product Usage: The C2 CryoBalloon Ablation System is intended for use as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications, to include ablation of Barrett s Esophagus with dysplasia. It is a cryosurgical system with a nitrous oxide cooled balloon that is compatible with a commercially available endoscope with a minimum working channel inner diameter (ID) of 3.7 mm and length of 100 cm. The C2 CryoBalloon System is a system comprised of a Catheter (sterile), Controller (non-sterile), and Cartridge (non-sterile). The Device is used to ablate unwanted tissue by application of extreme cold. The balloon at the distal end of the Catheter comes in contact with tissue and is inflated with nitrous oxide. Tissue is visualized through the inflated balloon, and the treatment site is selected by adjusting the endoscope and Controller position. The nitrous oxide spray cools the balloon to ablate the unwanted tissue, and the nitrous oxide exhausts through the Controller.

Reason

The Controller Cap of the C2 CyroBalloon Ablation System may crack as it is tightened onto the controller. That could compromise the integrity and result in damage to the Controller Cap.

Code Info

All lots distributed between June 23, 2016 and December 20, 2016: UPDATE - expanded to include product distributed from February 1 2016 through February 8, 2017. All numbers below have been updated. Lot numbers: Expiry dates 01122016-02 12-Jul-16; 06242016-01 1-Jan-2017; 11282016-03 2-Jun-2017; 01042016-03 14-Jul-16; 06292016-02 6-Jan-2017; 09202016-01 21-Mar-2017; 07062016-01 8-Jan-2017; 09222016-01 23-Mar-2017; 01252016-02 02-Aug-16; 07082016-02 13-Jan 2017; 09232016-01 28-Mar-2017; 01282016-01 01-Aug-16; 07082016-03 14-Jan-2017; 09232016-02 27-Mar-2017; 01292016-01 02-Aug-16; 07152016-01 21-Jan-2017; 09292016-01 4-Apr-2017; 02022016-03 09-Aug-16; 07152016-02 20-Jan-2017; 09302016-03 4-Apr-2017; 02082016-01 09-Aug-16; 07212016-01 25-Jan-2017; 10052016-02 19-Apr-2017; 02092016-02 11-Aug-16; 07252016-01 28-Jan-2017; 10142016-01 18-Apr-2017; 02112016-02 17-Aug-16; 07262016-01 9-Feb-2017; 10192016-01 24-Apr-2017; 02172016-02 25-Aug-16; 07282016-02 2-Feb-2017; 10212016-01 25-Apr-2017; 02222016-01 26-Aug-16; 08022016-01 17-Feb-2017; 10242016-01 26-Apr-2017; 03012016-01 04-Sep-16; 08032016-02 5-Feb-2017; 10262016-02 1-May-2017; 03072016-01 10-Sep-16; 08112016-01 16-Feb-2017; 10282016-02 8-May-2017; 03102016-02 15-Sep-16; 08232016-01 26-Feb-2017; 11022016-01 9-May-2017; 03152016-03 21-Sep-16; 08262016-01 2-Mar-2017; 11042016-01 10-May-2017; 03282016-02 29-Sep-16; 09022016-01 8-Mar-2017; 11092016-01 15-May-2017; 03292016-01 31-Sep-16; 09092016-01 14-Mar-2017; 11142016-02 16-May-2017; 09122016-02; 15-Mar-2017; 11152016-02 18-May-2017; 04012016-01 06-Oct-16; 09192016-01 19-Mar-2017; 11162016-01 21-May-2017; 09192016-02 21-Mar-2017; 11282016-02 30-May-2017; 04062016-01 12-Oct-16; 11302016-02 02-Jun-17; 12052016-01 08-Jun-17; 04112016-04 14-Oct-16; 12052016-03 06-Jun-17; 04142016-01 20-Oct-16; 12072016-01 09-Jun-17; 04202016-02 21-Oct-16; 12092016-01 13-Jun-17; 05022016-02 05-Nov-16; 12132016-01 19-Jun-17; 05092016-01 11-Nov-16; 12152016-02 20-Jun-17; 05162016-02 18-Nov-16; 01042017-01 04-Jul-17; 05232016-01 25-Nov-16; 01052017-03 19-Jul-17; 05272016-01 02-Dec-16; 01052017-04 19-Jul-17; 06022016-02 09-Dec-16; 01092017-01 12-Jul-17; 06102016-01 14-Dec-16; 01102017-01 18-Jul-17; 06152016-01 21-Dec-16; 01132017-01 18-Jul-17; 06172016-01 23-Dec-16; 01132017-03 17-Jul-17; 07282016-02 02-Feb-17.

Distribution

Worldwide Distribution - US Nationwide and the countries of Germany, Belgium, The Netherlands United Kingdom and China.

Quantity

755 controller caps