FDA Enforcement Class II Terminated

ANGIOGRAPHIC KIT W/MX123-3MR MANIFOLD, Model Number MX1030. for use during angiography and other special procedures.

Recall: Z-1475-2021 · Reported April 28, 2021

Enforcement

Recall Number
Z-1475-2021
Event ID
87647
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Smiths Medical ASD Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 28, 2021
Initiation Date
March 31, 2021
Classification Date
April 21, 2021
Termination Date
March 20, 2023
Address
6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States

Description

ANGIOGRAPHIC KIT W/MX123-3MR MANIFOLD, Model Number MX1030. for use during angiography and other special procedures.

Reason

Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.

Code Info

Lot Numbers: 3979813, 3982717, 4016454, 4024471

Distribution

Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.

Quantity

5540 devices