FDA Enforcement Class II Terminated

ZYM B Reagent (REF 70493). The ZYM B reagent is an additional test used to reveal the results of some miniaturized biochemical test included in the API strips. The API product line is a standardized system combining several biochemical tests, which enables group or species identifications of microorganisms.

Recall: Z-1475-2014 · Reported April 30, 2014

Enforcement

Recall Number
Z-1475-2014
Event ID
67820
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
BioMerieux SA
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 30, 2014
Initiation Date
March 6, 2014
Classification Date
April 18, 2014
Termination Date
March 14, 2017
Address
Chemin De L'Orme, N/A, Marcy L'Etoile, N/A, N/A, France

Description

ZYM B Reagent (REF 70493). The ZYM B reagent is an additional test used to reveal the results of some miniaturized biochemical test included in the API strips. The API product line is a standardized system combining several biochemical tests, which enables group or species identifications of microorganisms.

Reason

bioMerieux identified a visual defect and activity issue on the ZYM B reagent (REF 70493) and FB reagent (REF 70562). The activity issue lead to a false negative result of some Rapid ID 32 strips and API biochemical tests using the ZYM B reagent or FB reagent

Code Info

Codes - 1002706740, Exp. Date 10/14/2014, 1002717860, Exp. Date 10/24/2014, 1002742610, Exp. Date 10/30/2014, 1002760700, Exp. Date 11/4/2014, 1002821420, Exp. Date 11/26/2014, 1002857840, Exp. Date 12/12/2014, 1002860530, Exp. Date 12/12/2014

Distribution

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico, and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WV, WA, WI, and WY and the country of Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Benin, Bolivia, Bosnia, Brazil, Brunel, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Central African Republic, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Erythrea, France, Gabon, Georgia, Germany, Ghana, Greece, Guatemala, Haiti, Herzegovina, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jamaica, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Mali, Mexico, Moldavia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Thailand, Trinidad & Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.

Quantity

33,853