FDA Enforcement
Class II
Terminated
Gang Cath Lab Set, Model Number M20858R1.for the administration of IV fluids. Stopcocks and manifolds are used to control the direction of IV fluid flow.
Recall: Z-1474-2021
·
Reported April 28, 2021
Enforcement
- Recall Number
- Z-1474-2021
- Event ID
- 87647
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Smiths Medical ASD Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 28, 2021
- Initiation Date
- March 31, 2021
- Classification Date
- April 21, 2021
- Termination Date
- March 20, 2023
- Address
- 6000 Nathan Ln N, N/A, Minneapolis, MN, 55442-1690, United States
Description
Gang Cath Lab Set, Model Number M20858R1.for the administration of IV fluids. Stopcocks and manifolds are used to control the direction of IV fluid flow.
Reason
Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.
Code Info
Lot Numbers: 4007165
Distribution
Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.
Quantity
400 devices