FDA Enforcement Class II Ongoing

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Recall: Z-1472-2026 · Reported March 4, 2026

Enforcement

Recall Number
Z-1472-2026
Event ID
98332
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Agiliti Health - Ellis
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
March 4, 2026
Initiation Date
January 21, 2026
Classification Date
February 26, 2026
Address
204 W 2nd St, N/A, Ellis, KS, 67637-2026, United States

Description

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Reason

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Code Info

UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025

Distribution

US Nationwide distribution.

Quantity

4286