FDA Enforcement Class II Ongoing

Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower

Recall: Z-1471-2023 · Reported May 3, 2023

Enforcement

Recall Number
Z-1471-2023
Event ID
91938
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Draeger Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 3, 2023
Initiation Date
April 5, 2023
Classification Date
April 27, 2023
Address
6 Tech Dr, N/A, Andover, MA, 01810-2434, United States

Description

Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower

Reason

The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.

Code Info

UDI-DI: 04049098001878; All Systems Running Software versions: ICS VG2.1.3 and lower

Distribution

Worldwide - US Nationwide distribution.

Quantity

6803 systems