FDA Enforcement
Class II
Ongoing
Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower
Recall: Z-1471-2023
·
Reported May 3, 2023
Enforcement
- Recall Number
- Z-1471-2023
- Event ID
- 91938
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Draeger Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 3, 2023
- Initiation Date
- April 5, 2023
- Classification Date
- April 27, 2023
- Address
- 6 Tech Dr, N/A, Andover, MA, 01810-2434, United States
Description
Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower
Reason
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.
Code Info
UDI-DI: 04049098001878; All Systems Running Software versions: ICS VG2.1.3 and lower
Distribution
Worldwide - US Nationwide distribution.
Quantity
6803 systems