FDA Enforcement Class II Terminated

VITROS Chemistry Products Li (Lithium) Slides, REF/Catalog Number/Product Code 1632660 (18 slide format, 90 slides per sales unit) and REF/Catalog Number/Product Code 8318925 (60 slide format / 300 slides per sales unit), Unique Identifier Nos. 10758750009466 and 10758750004720, RX ONLY, IVD --Ortho Clinical Diagnostics, Inc. Rochester, NY 14626 --- Background: The VITROS Li Slide is a multilayered, analytical element coated on a polyester support. During the coating process, thin layers of reagents are coated onto a polyester support layer to create a spreading layer and a chemical (reagent) layer. During the coating process rolls of materials are produced. These rolls are segmented into slits. There are several slits in each roll. Each slit is then cut into small chemistry chips that are positioned between plastic to create the final Microslide format. Eighteen (18) or sixty (60) slides are placed into each plastic cartridge for use. Each slide is a device.

Recall: Z-1471-2017 · Reported March 22, 2017

Enforcement

Recall Number
Z-1471-2017
Event ID
76438
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 22, 2017
Initiation Date
February 1, 2017
Classification Date
March 15, 2017
Termination Date
June 29, 2018
Address
100 Indigo Creek Dr, N/A, Rochester, NY, 14626-5101, United States

Description

VITROS Chemistry Products Li (Lithium) Slides, REF/Catalog Number/Product Code 1632660 (18 slide format, 90 slides per sales unit) and REF/Catalog Number/Product Code 8318925 (60 slide format / 300 slides per sales unit), Unique Identifier Nos. 10758750009466 and 10758750004720, RX ONLY, IVD --Ortho Clinical Diagnostics, Inc. Rochester, NY 14626 --- Background: The VITROS Li Slide is a multilayered, analytical element coated on a polyester support. During the coating process, thin layers of reagents are coated onto a polyester support layer to create a spreading layer and a chemical (reagent) layer. During the coating process rolls of materials are produced. These rolls are segmented into slits. There are several slits in each roll. Each slit is then cut into small chemistry chips that are positioned between plastic to create the final Microslide format. Eighteen (18) or sixty (60) slides are placed into each plastic cartridge for use. Each slide is a device.

Reason

Intermittent imprecision in results generated using two specific lots of VITROS Li Slides.

Code Info

REF/Product Code 1632660: Lot 3124-0108-6568, exp. 2018- 04-01; REF/Product Code 8318925: Lot 3124-0108-6569, exp. 2018- 05-01

Distribution

US, Puerto Rico, Canada, EU, China, Brazil, Chile, and Dominican Republic.

Quantity

US: 94 sales units; OUS: 136 sales units