FDA Enforcement Class II Ongoing

TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number: 7209808

Recall: Z-1467-2022 · Reported August 3, 2022

Enforcement

Recall Number
Z-1467-2022
Event ID
90519
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Covidien Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 3, 2022
Initiation Date
June 7, 2022
Classification Date
July 26, 2022
Address
15 Hampshire St, Mansfield, MA, 02048-1113, United States

Description

TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number: 7209808

Reason

Electrical component failure within the control unit. The observed failure mode is a loss of function of the device when used with handpiece and foot switch.

Code Info

UDI-Device Identifier (GTIN/UPN) 10884521749931 Serial Number: YF1120, YF1126, YF1127, YF1128, YF1129, YF1130, YF1131, YF1133, YF1134, YF1135, YF1137, YF1138, YF1139, YF1140, YF1141, YF1142, YF1150, YF1152, YF1154, YF1156, YF1157, YF1158, YF1161, YF1163, YF1166, YF1167, YF1168, YF1169, YF1172, YF1175, YF1176, YF1177, YF1182

Distribution

AZ CA CO IL KS MA MI MN MO NY OK PA TN TX TX WA Foreign: United Kingdom Ireland Netherlands Reunion Spain

Quantity

33 units