FDA Enforcement Class II Ongoing

Medline procedure kits labeled as: 1) ACOUSTIC NEUROMA PACK-LF, Pack Number DYNJ37741B; 2) ANGIOGRAPHY PACK, Pack Number DYNJ21519W; 3) NEURO ANGIO PACK, Pack Number DYNJ83941; 4) NEURO ANGIO PACK SJF-LF, Pack Number DYNJ55678; 5) NEURO PACK, Pack Number DYNJ04851Q; 6) SNI NEURO, Pack Number DYNJ45379D; 7) VP SHUNT PACK, Pack Number DYNJ55558C

Recall: Z-1464-2025 · Reported April 2, 2025

Enforcement

Recall Number
Z-1464-2025
Event ID
96373
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
MEDLINE INDUSTRIES, LP - Northfield
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 2, 2025
Initiation Date
February 11, 2025
Classification Date
March 26, 2025
Address
3 Lakes Dr, N/A, Northfield, IL, 60093-2753, United States

Description

Medline procedure kits labeled as: 1) ACOUSTIC NEUROMA PACK-LF, Pack Number DYNJ37741B; 2) ANGIOGRAPHY PACK, Pack Number DYNJ21519W; 3) NEURO ANGIO PACK, Pack Number DYNJ83941; 4) NEURO ANGIO PACK SJF-LF, Pack Number DYNJ55678; 5) NEURO PACK, Pack Number DYNJ04851Q; 6) SNI NEURO, Pack Number DYNJ45379D; 7) VP SHUNT PACK, Pack Number DYNJ55558C

Reason

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.

Code Info

1) Pack Number DYNJ37741B: UDI/DI (each) 40193489682213 (case), Lot Number 20BDB113A; 2) Pack Number DYNJ21519W: UDI/DI (each) 40889942255611 (case), Lot Number 20BDB225A; 3) Pack Number DYNJ83941: UDI/DI (each) 40888277736574 (case), Lot Number 22JBA879Z; 4) Pack Number DYNJ55678: UDI/DI (each) 40193489468688 (case), Lot Number 22JBC551Z; 5) Pack Number DYNJ04851Q: UDI/DI (each) 40889942402176 (case), Lot Number 22JBC798Z; 6) Pack Number DYNJ45379D: UDI/DI (each) 40195327048748 (case), Lot Number 23FME560A; 7) Pack Number DYNJ55558C: UDI/DI (each) 40195327408214 (case), Lot Number 23LMG468A

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Quantity

422