nordicBrainEX, software versions 2.21 through 2.3.10
Recall: Z-1462-2022
·
Reported August 3, 2022
Enforcement
- Recall Number
- Z-1462-2022
- Event ID
- 90480
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- NordicNeuroLab AS
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 3, 2022
- Initiation Date
- June 16, 2022
- Classification Date
- July 25, 2022
- Termination Date
- September 4, 2024
- Address
- Mollendalsveien 65c, Bergen, N/A, Norway
Description
nordicBrainEX, software versions 2.21 through 2.3.10
Reason
When exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a multiframe dataset, the resulting output may have left/right sides flipped.
Code Info
UDI 07090042059015, software versions 2.21 through 2.3.10
Distribution
US Nationwide distribution.
Quantity
243 units