FDA Enforcement Class II Terminated

Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.

Recall: Z-1461-2018 · Reported April 25, 2018

Enforcement

Recall Number
Z-1461-2018
Event ID
79713
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cook Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 25, 2018
Initiation Date
March 22, 2018
Classification Date
April 19, 2018
Termination Date
May 13, 2020
Address
750 N Daniels Way, Bloomington, IN, 47404-9120, United States

Description

Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.

Reason

Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.

Code Info

All lots

Distribution

Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.

Quantity

2,267