FDA Enforcement
Class II
Terminated
Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607. Intended for the aspiration of body fluids and cells; in particular oocyte aspiration.
Recall: Z-1460-2018
·
Reported April 25, 2018
Enforcement
- Recall Number
- Z-1460-2018
- Event ID
- 79713
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Cook Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 25, 2018
- Initiation Date
- March 22, 2018
- Classification Date
- April 19, 2018
- Termination Date
- May 13, 2020
- Address
- 750 N Daniels Way, Bloomington, IN, 47404-9120, United States
Description
Cook Vacuum Pump (Cook Aspiration Unit) K-MAR-5200-US, GPN G51607. Intended for the aspiration of body fluids and cells; in particular oocyte aspiration.
Reason
Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
Code Info
All lots
Distribution
Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
Quantity
132