FDA Enforcement Class II Ongoing

Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT

Recall: Z-1458-2024 · Reported April 10, 2024

Enforcement

Recall Number
Z-1458-2024
Event ID
94128
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Covidien, LP
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 10, 2024
Initiation Date
February 14, 2024
Classification Date
April 1, 2024
Address
60 Middletown Ave, North Haven, CT, 06473-3908, United States

Description

Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT

Reason

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.

Code Info

Model Number: SIGSDS30CTVT; UDI/DI: 10884521741888; Lot numbers: N3L1909UY, N3L1868UY, N3L2008UY, N3L2221UY, N3M1541UY, N3M1603UY, N3M1639UY;

Distribution

Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.

Quantity

3,023 units