FDA Enforcement
Class I
Terminated
Ventstar Coax 230, MP03376
Recall: Z-1457-2023
·
Reported May 24, 2023
Enforcement
- Recall Number
- Z-1457-2023
- Event ID
- 92049
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Draeger Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 24, 2023
- Initiation Date
- April 17, 2023
- Classification Date
- May 17, 2023
- Termination Date
- September 3, 2024
- Address
- 3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States
Description
Ventstar Coax 230, MP03376
Reason
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Code Info
UDI-DI: 04048675545613; All lots.
Distribution
Domestic: US Nationwide Distribution.
Quantity
40 units