FDA Enforcement
Class II
Terminated
SOMATOM Sensation 64 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
Recall: Z-1456-2018
·
Reported May 9, 2018
Enforcement
- Recall Number
- Z-1456-2018
- Event ID
- 79428
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 9, 2018
- Initiation Date
- February 15, 2018
- Classification Date
- April 30, 2018
- Termination Date
- April 8, 2020
- Address
- 40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States
Description
SOMATOM Sensation 64 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
Reason
to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure
Code Info
8377520
Distribution
USA (nationally) Distribution.
Quantity
1301