FDA Enforcement Class II Ongoing

Centurion Manual surgical kits labeled as: a) HEGAR UTERINE DILATOR, 11-12MM, STERILE, Product Code HUD112ST; b) STERILE UTERINE DILATOR 1MM/2MM, Product Code HUD12ST; c) HEGAR UTERINE DILATOR, 5-6MM, STERILE, Product Code HUD56ST; d) HEGAR UTERINE DILATOR, 9-10MM, STERILE, Product Code HUD910ST; e) HANK UTERINE DILATOR 11" 7/8FR STERILE, Product Code I68920; f) HANK UTERINE DILATOR 11" 13/14FR STERILE, Product Code I68930

Recall: Z-1450-2024 · Reported April 10, 2024

Enforcement

Recall Number
Z-1450-2024
Event ID
94136
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
MEDLINE INDUSTRIES, LP - Northfield
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
April 10, 2024
Initiation Date
February 23, 2024
Classification Date
March 29, 2024
Address
3 Lakes Dr, Northfield, IL, 60093-2753, United States

Description

Centurion Manual surgical kits labeled as: a) HEGAR UTERINE DILATOR, 11-12MM, STERILE, Product Code HUD112ST; b) STERILE UTERINE DILATOR 1MM/2MM, Product Code HUD12ST; c) HEGAR UTERINE DILATOR, 5-6MM, STERILE, Product Code HUD56ST; d) HEGAR UTERINE DILATOR, 9-10MM, STERILE, Product Code HUD910ST; e) HANK UTERINE DILATOR 11" 7/8FR STERILE, Product Code I68920; f) HANK UTERINE DILATOR 11" 13/14FR STERILE, Product Code I68930

Reason

Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.

Code Info

a) HUD112ST, UDI/DI 50653160068288 (case) 00653160068283 (each), Lot Numbers: 2023032190; b) HUD12ST, UDI/DI 00653160288810 (case) 10653160288817 (each), Lot Numbers: 2023061690, 2023112990; c) HUD56ST, UDI/DI 50653160068257 (case) 00653160068252 (each), Lot Numbers: 2023112190; d) HUD910ST, UDI/DI 50653160068271 (case) 00653160068276 (each), Lot Numbers: 2023031390; e) I68920, UDI/DI 00653160351323 (case) 10653160351320 (each), Lot Numbers: 2023110290; f) I68930, UDI/DI 00653160351293 (case) 10653160351290 (each), Lot Numbers: 2023110990

Distribution

Worldwide distribution - US Nationwide and the countries of Panama, Canada.

Quantity

270 units