FDA Enforcement Class II Terminated

SOMATOM Scope Power Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Recall: Z-1450-2018 · Reported May 9, 2018

Enforcement

Recall Number
Z-1450-2018
Event ID
79428
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 9, 2018
Initiation Date
February 15, 2018
Classification Date
April 30, 2018
Termination Date
April 8, 2020
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

SOMATOM Scope Power Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.

Reason

to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure

Code Info

(Not Supplied)

Distribution

USA (nationally) Distribution.

Quantity

1301