FDA Enforcement Class II Terminated

Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.

Recall: Z-1449-2015 · Reported April 22, 2015

Enforcement

Recall Number
Z-1449-2015
Event ID
70847
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Hansen Medical Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
April 22, 2015
Initiation Date
March 18, 2015
Classification Date
April 15, 2015
Termination Date
September 23, 2015
Address
800 E Middlefield Rd, N/A, Mountain View, CA, 94043-4030, United States

Description

Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.

Reason

Based on investigation of two complaint incidents, all Hansen Medical Magellan Robotic System (Model No. 11132 and 11139) may experience an unexpected event during a retraction operation of catheter. Retraction of catheter may not stop when equipment button is released.

Code Info

All serial numbers.

Distribution

Worldwide Distribution - US Distribution and the countries of : Saudi Arabia, Australia, France, Italy, Germany, and United Kingdom.

Quantity

19 total installed systems