FDA Enforcement
Class II
Terminated
Pega Medical Inc., Fassier-Duval IM Telescopic System, male component (solid shaft), Part number: M040-SS-110; Used in Implant Catalogue number: FD-040(LS)-SS. orthopedic intramedullary rod.
Recall: Z-1448-2014
·
Reported April 16, 2014
Enforcement
- Recall Number
- Z-1448-2014
- Event ID
- 67675
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Pega Medical Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 16, 2014
- Initiation Date
- February 7, 2014
- Classification Date
- April 10, 2014
- Termination Date
- April 23, 2014
- Address
- 1111 Autoroute Chomedy, N/A, Laval, N/A, N/A, Canada
Description
Pega Medical Inc., Fassier-Duval IM Telescopic System, male component (solid shaft), Part number: M040-SS-110; Used in Implant Catalogue number: FD-040(LS)-SS. orthopedic intramedullary rod.
Reason
The Male Components in this lot are made of material with lower strength than manufacturing specifications.
Code Info
Lot #: 130424-15 (only laser marked on the Part Number: M040-SS-110); Lot #: 121109-11 &130424-15 [Indicated on the label of the Catalogue Number: FD-040(L)-SS].
Distribution
Worldwide Distribution - USA Nationwide in the states of Missouri, Texas, and the country of Germany.
Quantity
Total of 5 units (2 units in the US and 3 units in Germany)