FDA Enforcement Class II Terminated

Stryker Connected OR Cart,120 V Ref Cat. #240-099-155, Qty 1, Rx Only, Manufactured for: Stryker, 5900 Optical Court, San Jose, CA 95138, UDI: 07613327462784 - Product Usage: The Connect OR Cart is a mobile, medical-grade cart intended to store and transport components of minimally invasive and open-field imaging systems such as the Stryker Advanced Imaging Modality (AIM) System, PINPOINT Endoscopic Fluorescence Imaging System, or SPY Portable Handled Imaging (SPY-PHI) System) during intraoperative use. The Cart also provides power, video, and data connections to the components.

Recall: Z-1445-2020 · Reported March 11, 2020

Enforcement

Recall Number
Z-1445-2020
Event ID
84715
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 11, 2020
Initiation Date
April 23, 2019
Classification Date
March 5, 2020
Termination Date
June 3, 2021
Address
5900 Optical Ct, N/A, San Jose, CA, 95138-1400, United States

Description

Stryker Connected OR Cart,120 V Ref Cat. #240-099-155, Qty 1, Rx Only, Manufactured for: Stryker, 5900 Optical Court, San Jose, CA 95138, UDI: 07613327462784 - Product Usage: The Connect OR Cart is a mobile, medical-grade cart intended to store and transport components of minimally invasive and open-field imaging systems such as the Stryker Advanced Imaging Modality (AIM) System, PINPOINT Endoscopic Fluorescence Imaging System, or SPY Portable Handled Imaging (SPY-PHI) System) during intraoperative use. The Cart also provides power, video, and data connections to the components.

Reason

There is a potential problem with firm s Connected OR Cart, 120V resulting in cart monitor not attaching flush with their monitor stow handle. The product is not manufactured to specifications and is positioned on the incorrect side that may lead to monitor falling from the monitor arm and a risk of user injury.

Code Info

UDI: 07613327462784 Lot Numbers: 2019AN00005, 2019AN 00007, 2019AN00017, 2019AN00039, 2019AN00041, 2019AN00046, 2019AN00048, 2019AN00054, 2019AN00061, 2019AN00067, 2019AN00085, 2019AN00087

Distribution

Worldwide distribution - US Nationwide distribution in the states of GA, TX and countries of Canada, Singapore, Thailand, Malaysia. No US Govt.

Quantity

12 Devices