FDA Enforcement Class III Terminated

SoundWave Professional Suite Software 3.2, Ref CI-6055-014, SPN 058-023-32085, UDI: (01)07630016853227, Logical SoundWave Professional Suite 3.2, CE 0123 - Product Usage: Soundwave 3.2 is used by healthcare professionals to program Advanced Bionics sound processors for use with Advanced Bionics cochlear implants, a device is intended to restore a level of auditory sensation to children with profound sensorineural deafness via electrical stimulation of the auditory nerve.

Recall: Z-1441-2020 · Reported March 11, 2020

Enforcement

Recall Number
Z-1441-2020
Event ID
84800
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Advanced Bionics, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 11, 2020
Initiation Date
April 4, 2019
Classification Date
March 4, 2020
Termination Date
June 27, 2022
Address
28515 Westinghouse Pl, N/A, Valencia, CA, 91355-1398, United States

Description

SoundWave Professional Suite Software 3.2, Ref CI-6055-014, SPN 058-023-32085, UDI: (01)07630016853227, Logical SoundWave Professional Suite 3.2, CE 0123 - Product Usage: Soundwave 3.2 is used by healthcare professionals to program Advanced Bionics sound processors for use with Advanced Bionics cochlear implants, a device is intended to restore a level of auditory sensation to children with profound sensorineural deafness via electrical stimulation of the auditory nerve.

Reason

The manufacturer received complaints that customers were attempting to install the fitting software using the supplied USB drive, an error message was received, and installation failed. The failed installation has not impact to the current version of the software, and users are able to continue programming devices with the current software version.

Code Info

Lot #s: 663183 663188 663190 663858 663185 663184 663187 663186 664783 663860 664784 663859 669701 669100 669702

Distribution

US Nationwide distribution in the sates of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, OH, OK, PA, SC, SD, TN, TX, VA, VT, WI

Quantity

364 versions of software