Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Recall: Z-1440-2014
·
Reported April 16, 2014
Enforcement
- Recall Number
- Z-1440-2014
- Event ID
- 67677
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Draeger Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 16, 2014
- Initiation Date
- March 5, 2014
- Classification Date
- April 10, 2014
- Termination Date
- October 31, 2014
- Address
- 3135 Quarry Rd, N/A, Telford, PA, 18969-1042, United States
Description
Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Reason
The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.
Code Info
Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003, ASEK-0004, ASEK-0005
Distribution
USA Nationwide Distribution in the states of WI and NY.
Quantity
3