FDA Enforcement Class II Terminated

bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor controlled inspiratory valve.

Recall: Z-1439-2020 · Reported March 11, 2020

Enforcement

Recall Number
Z-1439-2020
Event ID
84337
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Vyaire Medical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
March 11, 2020
Initiation Date
November 25, 2019
Classification Date
March 4, 2020
Termination Date
August 8, 2023
Address
26125 N Riverwoods Blvd, N/A, Mettawa, IL, 60045-3420, United States

Description

bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor controlled inspiratory valve.

Reason

The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.

Code Info

Serial numbers: MB170104, MB170118, MB170151, MB170202, MB230000, MB230020, MB230024, MB230027, MB230031, MB230033, MB230035, MB230036, MB230037, MB230038, MB230039, MB230040, MB230041, MB230046, MB230048, MB230049, MB230052, MB230053, MB230054, MB230055, MB230056, MB230057, MB230058, MB230059, MB230060, MB230061, MB230062, MB230063, MB230064, MB230065, MB230066, MB230067, MB230068, MB230069, MB230070, MB230071, MB230072, MB230073, MB230074, MB230075, MB230076, MB230077, MB230078, MB230079, MB230080, MB230081, MB230082, MB230083, MB230084, MB230085, MB230086, MB230087, MB230088, MB230089, MB230090, MB230091, MB230092, MB230093, MB230094, MB230095, MB230097, MB230098, MB230099, MB230100, MB230101, MB230102, MB230104, MB230190, MB230191, MB230194, MB230195, MB230196, MB230197, MB230201, MB230204, MB230205, MB230207, MB230209, MB230210, MB230217, MB230219, MB230220, MB230221, MB230222, MB230227, MB230228, MB230229, MB230230, MB230231, MB230232, MB230233, MB230243, MB230244, MB230246, MB230247, MB230249, MB230251, MB230253, MB230256, MB230258, MB230260, MB230261, MB230262, MB230263, MB230266, MB230270, MB230274, MB230275, MB230277, MB230294, MB230297, MB230298.

Distribution

US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, MD, MI, PA, TX, WA and Puerto Rico.

Quantity

116